Section 3-7 - Completing the Hardware Installation
3-7-7
Switching the System ON/OFF
3-7-7-1
Switching the System ON
1) Verify that the ultrasound unit has been connected to the power supply and that the circuit breaker
is ON (if the system is being used on the SafeLock Cart), as described in the
Ultrasound Unit to the Electrical Outlet
2) Hold down the
On/Off
button on the control panel for 3 seconds. The system automatically performs
an initialization sequence which includes the following:
•
Loading the operating system.
•
Running a quick diagnostic check of the system.
•
Detecting connected probes.
3) When prompted, log in to the system as appropriate.
(Refer to the instructions for system login and user security setup as described in the
Vivid
i
n/Vivid q N User Manual).
The system first enters 2D-Mode with the probe and application that were last used before the system
was shut down. If the probe has been removed since the system was last used, the currently-connected
probes and their available applications are displayed and selected by default.
3-7-7-2
Switching the System OFF
NOTE:
After turning OFF the system, wait at least 10 seconds before turning it on again. The system may not
be able to boot-up if power is recycled too quickly.
The system can be switched OFF in one of three ways, as follows:
•
Holding down the
On/Off
button for 3-5 seconds will evoke the display of the System Exit dialog
box (see
). From this, you can either select
Full Shutdown
(click the
Shutdown
button)
or choose to place the system in
Standby Mode
(click the
Standby
button).
•
Holding down the
On/Off
button for more than 10 secs will cause the system to perform
Emergency
Shutdown.
This is not recommended under normal circumstances.
•
Closing the lid will place the system in
Standby Mode
.
Note:
The system will remain in
Standby Mode
for as long as battery power will allow.
NOTICE
When AC power is applied to the scanner, the
On/Off
button on the control console illuminates amber,
indicating the Back-end Processor is in
Standby
mode.
Figure 3-61 System Exit Dialog Box