g
GE Vingmed Ultrasound
Monitor Replacement Complete, S5 / V5
1. Introduction
1. Purpose
This document describes how to replace the monitor for System FiVe / Vivid FiVe.
2. Contents
The “Monitor Replacement Complete, System FiVe / Vivid FiVe” FC200267 consists of the
following parts:
FA200507 Cable, Monitor GND
FC307118 EIZO cable lock
FC307103 Adapter plate f. EIZO
FA307626 Lock for EIZO
066E0076 Monitor 17” EIZO
080A359B Plate screw 3.5x9.5
080B9040 Lock-nut M4 Hexagon
080B1060 Nut lock M6 Durlok
080C2066 Washer, flat (4 ea.)
080X3147 Plug, Skiffy (4 ea.)
FA314396 Label, System FiVe
FB314558 Label, Vivid FiVe
3. Tools
Tools needed for assembling:
Philips Screw driver
7mm and 10mm wrenches
Hammer
4. Assembling of the mechanical parts
CAUTION!
Please be aware of that the monitor weight is approx. 19 kg.
Two persons are needed when lifting the monitor.
10 December 2002