TO MEASURE ELECTRICAL SIGNALS
Millivolts
Select mV and DIRECT in the measure menu and press EXE to open
the working screen. The upper window in the working screen
indicates the measured millivolts reading. The range is 0-600 mV
split into two (auto) ranges of 0 to 100 mV and 100 to 600 mV. Read
the section “Scaled Readings” to use the SCALE option.
Volts
Select Volts and DIRECT in the measure menu and press EXE to
open the working screen. The upper window in the working screen
indicates the measured Volts reading. The range is 0-30 Volts split
into two (auto)ranges of 0 to 6 Volts and 6 to 30 Volts. Read the
section “Scaled Readings” to use the SCALE option.
Milliamps/XMT
Select mA and DIRECT in the measure menu and press EXE to open
the working screen. The upper window in the working screen
indicates the measured milliamps reading. The percentage of
4-20 mA is indicated in small size digits to facilitate easy alarm switch
adjustments. The range is 0-52 mA. Read the section “Scaled
Readings” to use the option for FLOW and LINEAR scaling.
Note:
XMT not applicable
to TRX-II iS