6 - [EN] English
K346 Issue 2
Operation
This section gives examples of how to connect and use the
instrument. Before you start:
• Read and understand the “Safety” section.
• Do not use a damaged instrument.
Operation - Thermocouple connections
Attach the thermocouple wires to the applicable
thermocouple mini-connector (Figure 1). The wider blade is
the negative.
Then attach the connector to the instrument.
Operation - Communications port connections
Use the communications port (A1 - item [9]) to attach an
IDOS Universal Measurement Module (UMM).
When you attach the cable from a UMM (Figure 7/8), the
instrument automatically changes the menus to give you
all the applicable options (Table 2/3).
Operation - Measure thermocouple values
To measure the temperature or mV values of a
thermocouple:
1. Connect the instrument (Figure 1) and, if necessary,
adjust the
Set Up
2. Select a
T1
or
mV
input task from
Select Task
and, if necessary, adjust the
Settings
Figure 1: Example configuration - To measure the
temperature or mV values of a thermocouple
Operation - Change the output values
When the output operation is set up (Table 5), use one of
these procedures to change the output values:
Table 7: Procedures to change the output
Example procedure (“Nudge” output):
Settings
Settings
°C
K
a) Temperature
b) mV
Output
Procedure
Select
Edit
(
■
■
) and/or use the
▲
▼
buttons.
See the example below.
,
Select
Start/Stop
(
■
■
) or use the
▲
▼
buttons
to make the step changes manually.
Select
Start/Stop
(
■
■
).
Display:
T1 output
1
Edit
2
3
Edit
4
■
■
■
■
▲
▼
+0100.0
Sk1 = Edit
Sk2 = Settings
Display:
T1 output
5
6
▲
▼
101.0
▲
▼
102.0
Increment = 1 (Table 5)