Chapter 4. Measuring and Recording Data
100
DMS Go Operator’s Manual
4.3
Displaying Measured Results (MEASURE- RESULTS)
The instrument is capable of displaying up to six measured readings at one time. The displayed readings are selected using the
RESULTS
submenu. The parameters available for display, which depend on the instrument configuration and the evaluation mode, include the
following:
•
LOC - The location of the reading (for example, A1 or B2)
•
GAIN - The gain value automatically set up by the DMS Go
•
M GAIN - The maximum gain value, entered by the user in the Measure menu.
•
TEMP - The current temperature of the material, entered by the user in the Probe & Cal menu.
•
THK - The thickness of the material measured by the DMS Go (not compensated by temperature)
•
T-THK - Temperature Compensated Thickness, or the thickness of the material as measured by the DMS Go, compensated by the
temperature
•
P-DLY - The probe delay, or the time between when the DMS Go generates a pulse and the time it leaves the probe and enters the
material (or its travel time through the probe, typically measured in microseconds)
•
C-THK - Coating thickness when in TopCOAT mode.
•