General System Requirements
1.3 Delivery Requirements
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
1-11
Optional Items
NM UPS
May vary but not
more than
60×40×80
May vary but
not more than
150
any
any
any
any
any
May vary but
not more
than 130
ECG Trigger Monitor
May vary but not
more than
80×80×80
May vary but
not more than
15
any
any
any
any
any
<13
Xeleris
Monitors
NM600 Detectors Dismount Option
NM gantry without detectors
220×150×168
2125
94.5
94.5
222
188
195
**
1640
††
(with dollies)
Detector 1
93×86×100
320
86
86
100
100
98
‡‡
320
***
Detector 2
93×86×100
320
86
86
100
100
*
The minimum door width required in order to bring the system components into the scan room also depends on the width of the corridor leading
to the room. When planning or measuring the width of the scan room door, use the graphs provided in
Figure 1-2: D670 Relative Required Width
for Corridor and Scan Room Door, p.1-17
in order to verify that the measurements comply with the requirements.
†
The corridor width required in order to move the system components from the unloading area to the scan room depends on the
angles of turns on the corridor. See
Figure 1-3: D670 Required Corridor Width for 90° Turns, p.1-18
for the required width when the angle is 90°.
‡
May be delivered a few days prior to system delivery, as part of the final room check and preparation for installation.
** 20 mm clearance above the floor
†† Weight of gantry in the in-site transportation configuration: 1330 kg + weight of the dolly: 310 kg
‡‡ 50 mm clearance above the floor
*** The specified weight includes the packing. The detectors must be conveyed crated (unpacking is allowed only at the room or designated area
where the detectors to be installed)
Table 1-1: D670 Components and Clearance — Metric(Continued)
Crated
Uncrated
Component name
Crate size
(Dollies not attached)
(cm) (Height × Width ×
Depth/Length)
Weight
(kg)
Minimal dimensions
(cm)
*
Weight
(kg)
Door
width
Corridor /
elevator
width
Corridor /
elevator
length
Width of
corridors with
90° turns
†
Height