General System Requirements
1.2 Customer Responsibilities
1.2.1 Using Radioactive Isotopes
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
1-3
Complying with all national, state, or local regulatory requirements for the country in which
the installation occurs, for example:
Fire control devices as required by local codes
Permits, inspections, radiation licensing etc.
Earthquake-related regulations
Local regulations for Service Clearance & Egress
Assuring regulatory compliance for the use of radioactive isotopes and preparation of the
required isotopes (see
Using Radioactive Isotopes, p.1-3
)
Safe storage of the system and auxiliary equipment prior to and during installation
Floor tile removal and replacement in area of table and gantry
Ensuring adequate accessibility for all system components and auxiliary equipment to the site
1.2.1 Using Radioactive Isotopes
Since the system involves the use of radioactive isotopes, compliance with Nuclear Regulatory
Commission regulations, or similar regulatory requirements (depending on the country), must be
adhered to and all permissions obtained well in advance. It is recommended that regulatory
compliance is arranged early in the site planning process.
It is essential that all preparations are completed so that required source materials can be obtained
prior to installation, including calibration sources. Take into consideration that these sources may
have fairly long delivery lead times, yet may also have a short half life, so that it may not be
advisable to store them over long periods of time.