Page 248
Chapter 8 - System QC, Calibration and Functional Tests
8.1.1.5 VQC Processing and Software Realignment
1.) Set the system into PET mode by clicking the [
PET
] icon on the Image Monitor desktop area.
The Scan Monitor displays the PET main screen.
2.) Click the [
Full QA
] button on the PET main screen.
3.) Select
Volumetric QC
from the drop-down menu. The Volumetric QC panel shown in
displays.
Figure 8-6:
Initial Volumetric QC Panel
. Click [
Load
] to load the image sets that represent the CT and PET planes.
The
Images Database Navigator
displays showing the Patient List.
Figure 8-7:
Images Database Navigator
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......