6.2.7.9 Attach the Roller Cover Canister to the Secondary Base
For best results, use two people to install the Roller Cover Canister, P/N 2331848.
Use a 5 mm Allen wrench to fasten the canister assembly to the table with four sets of the following
M6 hardware:
• M6 x 16 socket-head cap screw, P/N 46-328417P13
• M6 flat washer, P/N 46-328430P3
• M6 lock washer, P/N 46-328432P3
1.) Refer to
. Flip the canister assembly forward, so the drip tray and cover face the
ceiling.
Carefully
lower the canister assembly between the gantry and table.
2.) Refer to
. Each canister bracket assembly arrives with two of the M6 screws
fastened through the bottom of the bracket. Slide the lower set of canister holes over the
screws to support the weight while aligning and fastening the assembly into place.
3.) Center the canister to the secondary base cover guides, and check for levelness before
fastening it into place.
Figure 6-122:
Roller Canister in Place (P/N 2331848)
Each bracket arrives with bottom screw in place.
M6 Hardware
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......