GE
HEALTHCARE
D
IRECTION
2307224-100, R
EVISION
6
D
ISCOVERY
LS S
YSTEM
S
ERVICE
M
ANUAL
Chapter 6 - Table
Page 177
6.2.6.35 Install Optical Sensor (P/N 2350570)
CAUTION
Refer to
. To prevent damage to the optical sensor, temporarily fasten it into
place with one screw. After the sensor mounting bracket is adjusted and the clearance
checked, fasten the sensor into place with both screws.
Figure 6-93:
Fasten Optical Sensor with One Screw
1.) Refer to
. Use a 3 mm Allen wrench to attach the optical sensor to the CT-PET
Position Latch with a single M3 x 10 mm screw, P/N 46-214695P7.
2.) Refer to
. Pivot the optical sensor away from the table rail.
3.)
Slowly
move the table into the PET detent.
4.) Refer to
Carefully
swing the optical sensor over the flag.
Figure 6-94:
Swing the Optical Sensor into Place
Bend this bracket to adjust
the optical sensor with
respect to the flag plate.
Flag Plate
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......