Page 172
Chapter 6 - Table
6.2.6.32 Install the PET Detent and Flag Assembly
The PET Detent/Flag assembly attaches near the middle of the right table rail with three M5
mounting bolts. Center the bolts in the slots when fastening the PET Detent/Flag assembly to the
rail. Refer to
. The PET detent has slots to provide Z-Axis adjustment.
Figure 6-86:
PET Detent/Flag Assembly
1.) Use a 4 mm Allen wrench to temporarily fasten the PET detent to the right table rail.
•
All table rail revisions:
Use three M5 x 12 mm Bolts, P/N 46-214695P23, to fasten the PET
detent to the right table rail.
• Center the hardware in the slots, and snug into place.
Flag
Center the hardware in the slots.
Flag Adjuster
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......