6.2.6.28 Install the CT and PET Detent/Flag Assemblies
The CT-PET Position Latch rolls into two detents that attach to the right side rail. The first time, the
detents are attached directly to the rail. During the latch adjustments, one or more shims may need
to be inserted behind the detents. This section contains the instructions to install the CT and PET
detents, including the side rail access plugs.
Refer to
. If necessary, measure the distance from the rear edge of the right side rail to
locate the detent mounting holes.
Figure 6-80:
CT and PET Detent/Flag Assembly Locations
PET Detent Location
CT Detent Location
(1,261 mm from end of secondary
(598 mm from end of secondary
base rail at rear of table to rear
mounting hole.)
Measure from the back edge of
base rail to rear mounting hole.)
rail to first (rear) mounting hole.
Содержание Discovery LS
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION 2307224 100 Rev 6 Discovery LS System Service Manual ...
Страница 2: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 2 ...
Страница 24: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 24 Table of Contents ...
Страница 32: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 32 List of Figures ...
Страница 110: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 110 Chapter 5 Console ...
Страница 222: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 222 Chapter 6 Table ...
Страница 242: ...GE HEALTHCARE DIRECTION 2307224 100 REVISION 6 DISCOVERY LS SYSTEM SERVICE MANUAL Page 242 Chapter 7 Gantry ...
Страница 357: ......