18
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
Revision A
2018554-001
6. Using the inflation bulb, apply the following pressures (measured
by an external digital manometer) and confirm that the Monitor
readings agree with the following table for both PT1 and PT2
channels.
7. If calibration is required, refer to Section 4.7.1.
1.6.4 Neonate Overpressure Check
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select
More...
3. Select the
Service
mode and input
2 2 1 3
.
4. From the Service menu, select the
NIBP
button, then select the
Popoff
button.
Note:
If the overpressure point is out of range, adjust the
overpressure potentiometer as described in 4.7.2.
5. Using the inflation bulb, increase applied pressure until
overpressure occurs. Confirm that pressure at that point is between
150 mmHg to 165 mmHg and system pressure falls to less than 20
mmHg within 4 seconds.
6. Power off the PRO Monitor.
7. If calibration is required, refer to Section 4.7.1.
1.6.5 Adult Overpressure Check
1. Set up the unit and calibration equipment as shown in Figure 4-1.
2. Power on the PRO Monitor and select
More...
3. Select the
Service
mode and input
2 2 1 3
.
Содержание Dinamap PRO Series
Страница 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410...
Страница 2: ......
Страница 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Страница 35: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual 27 2018554 001 1 11 ERROR CODES...
Страница 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...
Страница 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix...
Страница 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001...
Страница 41: ......