DigitalFlow™ GS868 Service Manual
47
Appendix B. Optional Enclosures
B.4 Rack Mount Option Card Installation
The Model GS868 flowmeter can accommodate up to six option cards in a manner similar to that used in a PC. The
option cards are installed into sockets on the printed circuit board, and they are held in place with a metal bracket. A
single metal bracket is used to secure all the installed option cards.
Note:
If the Model GS868 presently has no option cards installed, be sure to purchase the metal mounting bracket
along with the first option card.
Refer to Figure 11 on page 54 and complete the following steps:
1.
After disconnecting the main power to the meter, remove the
top panel
from the enclosure by removing the four
screws indicated.
2.
If one or more option cards are already installed, remove the four
fasteners
that secure the metal bracket to the
printed circuit board. Lift the
metal bracket
straight up and away from the printed circuit board.
Note:
The fasteners may be either plastic snap rivets or metal hardware (in some older meters). In either case, they
may be discarded, as new plastic snap rivets will be provided.
3.
There are six 32-pin
option card sockets
(
J41–J46
) on the rear of the printed circuit board. By removing the screw
on the rear panel, remove the appropriate
option slot cover
to make room for the external connector on the new
option card.
4.
To install the
option card
, insert its 32-pin connector into the chosen option card socket and gently press the card
into place. Make sure that the pins in the connector are straight and properly aligned with the socket and that the
option card is oriented with its external connector positioned in the opening on the rear panel.
5.
Repeat steps 3 and 4 to install any additional option cards.
6.
Place the metal bracket over the option cards, making sure that all installed option cards are inserted into the plastic
card guides in the bracket. The bracket must be oriented so that the six card guides are directly above the six
sockets on the printed circuit board (
do not
install the bracket rotated 180° out of position). Secure the metal
bracket to the printed circuit board with the four plastic snap rivets provided.
7.
After checking for any loose hardware that may have fallen into the enclosure, reinstall the top panel on the meter
and secure the panel in place with the four screws previously removed.
The Model GS868 flowmeter may now be placed back into service. Reconnect the main power and refer to Appendix
C,
Optional Enclosures
, of the
Startup Guide
for instructions on wiring the new option card(s).
Note:
Be sure to record the option card installation in Appendix A,
Service Record
.
WARNING!
The main power to the Model GS868 must be disconnected before proceeding.
CAUTION!
Do not force the option card into the socket. If the card does not enter the socket
easily, check for and straighten any bent pins in the connector and try again.
Содержание DigitalFlow GF868
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 10: ...Contents viii DigitalFlow GS868 Service Manual ...
Страница 26: ...Chapter 1 Calibration DigitalFlow GS868 Service Manual 16 ...
Страница 40: ...Chapter 3 Diagnostics 30 DigitalFlow GS868 Service Manual no content intended for this page ...
Страница 54: ...Appendix A Service Record 44 DigitalFlow GS868 Service Manual no content intended for this page ...
Страница 62: ...Appendix B Optional Enclosures 52 DigitalFlow GS868 Service Manual no content intended for this page ...
Страница 68: ...Warranty 58 DigitalFlow GS868 Service Manual no content intended for this page ...
Страница 72: ......
Страница 73: ......