Calibration
1-13
April 2008
Testing the Totalizer/
Frequency Outputs
Totalizer/frequency outputs may be added to the Model GC868 by
installing a
Totalizer/Frequency Option Card
in one (or more) of the
six expansion slots. Each option card includes four outputs, which are
designated as A, B, C and D. To test the outputs, connect a frequency
counter to the card’s terminal block as shown in Figure 1-5 below.
Refer to the menu map in Figure 1-6 on page 1-15 as a guide.
Figure 1-5: Frequency Counter Connections
For this discussion, assume that the option card has been installed in
Slot x
.
Accessing the Calibration
Menu
1.
Press the
[CAL]
key to enter the
Calibration Program
.
2.
Press
[Fx]
to calibrate the desired slot. (The option bar will include
a slot listing for each installed option card.)
3.
Press
[F1]-[F4]
to select output A, B, C or D, respectively.
Entering Frequency
4.
Enter a new frequency in the range of 1-10,000 Hz and press the
[ENT]
key. Verify that the frequency counter reads the correct
value.
Terminal Block (Option Card)
A
NO
A
COM
B
B
D
D
C
C
Frequency
1
8
Counter
IN
COM
COM
NO
COM
NO
COM
NO
Содержание DigitalFlow GC868
Страница 6: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 31: ...Chapter 3 ...
Страница 39: ...Chapter 4 ...
Страница 52: ...Appendix A ...
Страница 53: ...Service Record Introduction A 1 Data Entry A 1 Diagnostic Parameters A 3 ...
Страница 55: ...April 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 57: ...Appendix B ...