March 2008
B-2
Optional Enclosures
Rack Mount Fuse
Replacement
If it has been determined that the fuse in the Model DF868 requires
replacement, refer to Figure B-1 on page B-10 and Figure B-2 on
page B-11, and complete the following steps:
1.
On the rear panel of the meter, set the power switch to the OFF
position and pull the power cord out of its receptacle.
!WARNING!
The main power to the Model DF868 must be
disconnected before proceeding.
2.
Locate the black plastic fuse cover on the back panel at the bottom
of the power cord receptacle and pull it straight out.
3.
Replace the defective fuse with a new one of the same rating and
type. Use only 5 x 20 mm Slo-Blo fuses, having a rating as
indicated in Table B-1 below and on the rear panel label.
4.
Reinstall the black plastic fuse cover and insert the power cord
back into the receptacle. Then, position the power switch in the
ON position.
The Model DF868 flowmeter may now be placed back into service.
Reconnect the main power and resume taking measurements.
Note:
Be sure to record the fuse replacement in Appendix A,
Service
Record
.
Table B-1: Line Voltages & Fuse Ratings
Line Voltage
Fuse Rating
100-120 VAC
1.0 A, Slo-Blo
220-240 VAC
0.5 A, Slo-Blo
12-28 VDC
3.0 A, Slo-Blo
Содержание DigitalFlow DF968
Страница 2: ......
Страница 7: ...Chapter 1 ...
Страница 23: ...Chapter 2 ...
Страница 32: ...Chapter 3 ...
Страница 42: ...Chapter 4 ...
Страница 55: ...Appendix A ...
Страница 56: ...Service Record Introduction A 1 Data Entry A 1 ...
Страница 58: ...March 2008 A 2 Service Record Table A 1 Service Record Continued Date Description of Service Performed By ...
Страница 60: ...Appendix B ...
Страница 72: ...March 2008 Optional Enclosures B 11 Figure B 2 Model DF868 Rack Mount Console Assembly ref dwg 705 849 ...