DigitalFlow™ XGM868i Service Manual
43
Appendix B. Calibration and Testing with PanaView
B.3 Calibrating Slot 0 Analog Outputs
Every Model XGM868i flowmeter includes two built-in analog outputs (A and B) at terminal block
J1
, which is
designated as
Slot 0
. Both the zero-point and full-scale values for each output must be calibrated. After calibrating the
outputs, which have a resolution of 5.0
A (0.03% full scale), their linearity should be tested.
Note:
The zero point of the analog output may be set for either 0 mA or 4 mA. However, the calibration procedure
always uses the 4 mA point, as the meter will extrapolate this value to obtain the 0 mA point.
Prepare for calibration by inserting an ammeter in series with the load on analog output A at terminal block J1, as
shown in Figure 11.
Figure 11: Ammeter Connection (Output A)
1.
From the
Calibration/Test
menu (see Figure 10 on page 42), double-click on the
Calibration
entry to open the
option.
2.
In the center pane, double-click on
Slot 0
.
Note:
The
Slot 1
option only appears in the center pane if an option card has been installed in this slot.
Terminal Block J1 (Slot 0)
B–
RTN
B+
SIG
A–
RTN
A+
SIG
4
1
+
-
Ammeter
Load
Содержание DigitalFlo XGM868i
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 44: ...Chapter 4 Parts Replacement DigitalFlow XGM868i Service Manual 36 ...
Страница 48: ...40 DigitalFlow XGM868i Service Manual no content intended for this page ...
Страница 70: ...62 DigitalFlow XGM868i Service Manual no content intended for this page ...
Страница 78: ...Warranty 70 DigitalFlow XGM868i Service Manual no content intended for this page ...
Страница 79: ......