10 Specifications
78
GE Healthcare
86500-IMG rev 3
10.1 E
LECTROMAGNETIC
C
OMPATIBILITY
(EMC)
TABLES
The Diamond is suitable for use in the specified electromagnetic environment. The purchaser or user
of the Diamond should assure that it is used in an electromagnetic environment as described below:
Emissions Test
Compliance
Electromagnetic Environment
Radio-Frequency
Emissions
CISPR11
Group 1
The Diamond uses RF energy only for its internal function.
Therefore, the RF emission is very low and not likely to cause
any interference in nearby electronic equipment.
Radio-Frequency
Emissions
CISPR11
Class B
The Diamond is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not
applicable
The Diamond is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not
applicable
The Diamond is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Table 10.1 Electromagnetic emissions IEC 60601-1-2
The Diamond is suitable for use in the specified electromagnetic environment. The purchaser or user
of the Diamond should assure that it is used in an electromagnetic environment as described below:
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
±
2, 4, 6 kV for
contact discharge
±
2, 4, 8 kV for air
discharge
±
2, 4, 6 kV for
contact discharge
±
2, 4, 8 kV for air
discharge
Floors are wood, concrete, or ceramic
tile, or floors are covered with
synthetic material and the relative
humidity is at least 30 percent.
Electrical fast
transient/burst
IEC 61000-4-4
±
2 kV for power
supply lines
±
1 kV for
input/output
lines
±
2 kV for power
supply lines
±
1 kV for
input/output
lines
Mains power quality is that of a
typical commercial and/or hospital
environment
Surge
IEC 61000-4-5
±
1 kV differential
mode
±
2 kV common
mode
±
1 kV differential
mode
±
2 kV common
mode
Mains power quality is that of a
typical commercial and/or hospital
environment.
Table 10.2 Electromagnetic immunity IEC 60601-1-2
FOR
TRAINING
PURPOSES