Installation
Mounting the MMY31
(cont.)
Mount the MMY31 using a compression fitting connection. Be
certain that the tip of the probe does not touch the inside wall of the
pipe.
IMPORTANT:
Ensure that your probe has a stainless steel ferrule,
which is required for use in pressurized systems.
Adjust the ferrule for a probe insertion length of 1” (minimum) and
tighten the compression fitting as follows (see Figure 2-2 below):
1.
Hand-tighten the nut.
2.
Then, using a wrench, tighten the nut 1½ turns. Now the fitting is
tight and can withstand pressure to 1750 psig (120 barg).
Figure 2-2: Mounting the MMY31
Minimum 1 in. (25 mm)