General System Information
1-3
November 2005
Introduction
Unit Description
The DewPro MMY30 trace moisture transmitter (shown in Figure
1-2 below) is a loop-powered dewpoint measuring device. The
transmitter includes a sensor element, a flow chamber, a weather-
proof enclosure, microprocessor electronics, and assorted fittings, all
in a compact assembly. In most cases, either the inlet or outlet port
includes an orifice to regulate the flow. The placement of this orifice
determines whether the dew point measurement is done at process
(line) pressure (outlet orifice), or at atmospheric pressure (inlet
orifice). A 2 micron sintered inlet filter prevents particles from
entering the device.
Optional Display/
User Interface
The optional display/user interface feature allows the DewPro to be
configured to the user's specifications. See Chapter 4 for more
information.
Figure 1-2: DewPro MMY30 Transmitter
Содержание DewPro MMY30
Страница 1: ...GE Sensing DewPro MMY30 General Eastern Trace Moisture Transmitter User s Manual ...
Страница 6: ...Chapter 1 ...
Страница 13: ...Chapter 2 ...
Страница 19: ...Chapter 3 ...
Страница 24: ...Chapter 4 ...
Страница 34: ...Chapter 5 ...
Страница 35: ...Troubleshooting Problems and Recommended Solutions 5 1 Removing the Filter 5 1 ...
Страница 37: ...Chapter 6 ...
Страница 38: ...Technical Specifications MMY30 Specifications 6 1 Optional Onboard Display with User Interface 6 2 ...
Страница 41: ...Appendix A ...
Страница 42: ...User Interface Matrix Input ...