Aisys
2-22
04/08 M1046983
2.8 Ventilator Interface board
The Ventilator Interface board (
A
) provides the electrical and/or pneumatic interface to
the following:
• Inspiratory (
B
) and expiratory (
C
) flow sensors (transducers)
• Patient airway (
D
) and manifold (
E
) pressures (transducers)
• Oxygen sensor (in breathing system)
• Inspiratory Flow Valve
• Gas Inlet Valve
• SCGO solenoid, SCGO/CGO position switches (if SCGO installed)
• Canister release switch
• ABS On switch
• ACGO position switch (if ACGO installed)
• O
2
Flush switch
• Bag/Vent switch
• Accessory Power (for task lights)
The Ventilator Interface board functions are managed locally by a microcontroller. The
microcontroller communicates data values to the controlling CPU via an RS-422 serial
interface.
Figure 2-17 • Ventilator Interface board
A
B
C
E
D
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......