Aisys
4-28
04/08 M1046983
4.5.10 Mixer P Zero
The Mixer P Zero instructions appear on the Mixer P Zero menu.
Instructions
Note
If repeated zero attempts fail, follow the procedure below:
1. Gain access to the components in the pan electrical enclosure
2. Disconnect the Alt O2 inlet tubing elbow fitting from the Mixer manifold.
3. Disconnect the tubing from the outlet elbow fitting.
4. Repeat the Mixer P Zero following the instructions as they appear on the
screen.
5. Reassemble in reverse order.
Back
Default
Start
To go back to factory defaults, select Defaults (above).
To Zero Pres Sensors:
1. Disconnect pipeline gas supplies.
2. Close the gas cylinders.
3. Remove the flow sensors.
4. Insert a non-Des cassette.
5. Push the O2 Flush button for 3 seconds.
6. Let the system sit WITHOUT gas flow for > 5 min.
7. Select Start.
Do not disturb the system while waiting for results.
Содержание Datex-Ohmeda Aisys Carestation
Страница 1: ...Aisys Anesthesia Machine Technical Reference Manual ...
Страница 16: ...Notes xiv 04 08 M1046983 ...
Страница 32: ...Notes 1 16 04 08 M1046983 ...
Страница 35: ...2 Theory of Operation M1046983 04 08 2 3 A M AB 75 098 L H D AB 75 046 J G I F K P O E D B C H N ...
Страница 158: ...Notes 5 20 04 08 M1046983 ...
Страница 288: ...Notes 7 106 04 08 M1046983 ...
Страница 325: ...9 Repair Procedures M1046983 04 08 9 29 The rear inverter A B C ...
Страница 356: ...Notes 9 60 04 08 M1046983 ...
Страница 397: ...10 Illustrated Parts M1046983 04 08 10 41 1 3 4 2 17 15 16 5 6 13 non DES cassette DES cassette 7 8 11 12 10 5 14 18 9 ...
Страница 399: ...10 Illustrated Parts M1046983 04 08 10 43 14 1 2 3 6 5 4 5 8 7 15 16 15 12 17 10 9 11 13 or 18 ...
Страница 401: ...10 Illustrated Parts M1046983 04 08 10 45 14 1 2 3 6 5 4 5 8 7 15 16 18 17 10 9 11 12 13 or ...
Страница 403: ...10 Illustrated Parts M1046983 04 08 10 47 14 15 18 1 2 3 6 5 4 5 8 7 17 10 9 11 15 16 12 13 or 17a 17c 17d 17b 17e ...
Страница 452: ...Notes 10 96 04 08 M1046983 ...
Страница 466: ...11 14 04 08 M1046983 Notes ...
Страница 470: ...Aisys 12 4 04 08 M1046983 12 3 System Schematics 12 3 1 Power Schematic ...
Страница 478: ...Aisys 12 12 04 08 M1046983 12 6 3 Tools Transfer Logs ...
Страница 495: ...12 Service Application M1046983 04 08 12 29 12 10 2 Vaporizer Subsystem Perform Vaporizer Test ...
Страница 497: ......