Field Replaceable Units and Upgrades:
Display Assembly
Replacing the Display Assembly
Dash 3000 Display Assembly
1. Remove the handle assembly. Refer to page 8-7.
2. Remove the display assembly from the main unit. Refer to page 8-9.
3. Replace the defective display assembly.
4. Reassemble the monitor in reverse order.
Dash 4000 with the Sharp Display Assembly
CAUTION
FLEX CABLE DAMAGE — DO NOT kink, pinch,
stretch, twist, or tightly fold a flex cable.
1. Remove the handle assembly. Refer to page 8-7.
2. Remove the display assembly from the main unit. Refer to page 8-9.
3. Remove the new display flex from the package and carefully install it
on the display assembly.
NOTE
You will need a thick book or some heavy object to support the
display assembly while you attach the display flex to the main
unit.
a. Insert the flex cable connector into the display connector as
shown in the figure below.
b. Using two of the flathead M3X6LG screws from the kit, secure
the flex cable to the display assembly.
873A
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......