8-4
Dash 3000/4000 Patient Monitor
Revision A
2000966-339
Field Replaceable Units and Upgrades:
Disassembly Guidelines
Hardware Assemblies
When disassembling the monitor, observe the following guidelines:
Remove the handle assembly, then remove the display assembly to
access the field replaceable or upgradeable components of the display
assembly and the main unit.
Note the positions of wires, cables, and different sized screws;
marking them if necessary to ensure they are replaced correctly.
DO NOT kink, pinch, stretch, twist, or tightly fold a flex cable.
PCB Assemblies
Electrostatic Discharge (ESD) Precautions
All external connector inputs and outputs of the monitor are designed
with protection from ESD damage. However, if the monitor requires
service, exposed components and assemblies contained within are
susceptible to ESD damage. This includes human hands, non-ESD
protected work stations and/or improperly grounded test equipment.
The following guidelines help make a service workstation more resistant
to the ESD damage:
Discharge any static charge you may have built up before handling
semiconductors or assemblies containing semiconductors.
A grounded, antistatic wristband (3M part number 2046 or
equivalent) or heel strap should be worn
at all times
while handling
or repairing assemblies containing semiconductors.
Use properly grounded soldering and test equipment.
Use a static-free work surface (3M part number 8210 or equivalent)
while handling or working on assemblies containing semiconductors.
DO NOT remove semiconductors or assemblies containing
semiconductors from antistatic containers (Velo-stat bags) until
absolutely necessary.
Make sure power to an assembly is turned off before removing or
inserting a semiconductor.
DO NOT slide semiconductors or electrical/electronic assemblies
across any surface.
DO NOT touch semiconductor leads unless absolutely necessary.
Semiconductors and electrical/electronic assemblies should be stored
only in antistatic bags or boxes.
Handle all PCB assemblies by their edges.
DO NOT flex or twist the circuit board.
These guidelines may not guaranty a 100% static-free workstation, but
can greatly reduce the potential for failure of any electrical/electronic
assemblies being serviced.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......