Revision A
Dash 3000/4000 Patient Monitor
4-55
2000966-339
Maintenance:
Checkout Procedures
Battery Tests
1. Disconnect the power cord plug from the wall receptacle.
2. Verify the
BATTERY
front panel indicator illuminates. This indicates
operation from the monitor’s battery power.
3. Setup the patient simulator as follows:
ECG heart rate – 80 bpm,
ECG amplitude – 1.0 mV,
5-lead patient cable attached.
4. Observe the following:
ECG Lead II is displayed and is noise-free,
Heart rate of 80 ±1 bpm is displayed,
With
QRS VOLUME
enabled, an audible tone sounds with each
R-Wave.
5. Verify all six ECG leads are selectable for display on the monitor.
6. Connect the power cord plug to the wall receptacle.
7. Verify the AC front panel indicator illuminates. This indicates the
monitor is operating from wall receptacle (AC) power.
8. Verify the
CHARGING STATUS
front panel indicator illuminates
for a few minutes.
An amber glow indicates the monitor battery is charging.
A green glow indicates the monitor batteries are fully charged.
Graph Test
Using the
Trim Knob
control, access the Service Mode menu starting
from the Main
Menu.
1. Select
MORE MENUS > MONITOR SETUP > SERVICE MODE.
2. Enter password using the
Trim Knob
control to select the day and
month from monitor screen with leading zeros. (e.g. July 4 = 0407).
3. Select
GRAPH TEST PATTERN > START
.
4. Verify the following:
Fonts.
Shading.
Triangle Pattern.
No missing dots.
5. Select
GRAPH TEST PATTERN > STOP.
Содержание Dash 3000
Страница 1: ...Dash 3000 4000 Patient Monitor Service Manual 2023896 023 Revision A ...
Страница 11: ...Revision A Dash 3000 4000 Patient Monitor 1 1 2000966 339 1 Introduction ...
Страница 12: ...1 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 19: ...Revision A Dash 3000 4000 Patient Monitor 2 1 2000966 339 2 Equipment Overview ...
Страница 20: ...2 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 45: ...Revision A Dash 3000 4000 Patient Monitor 3 1 2000966 339 3 Installation ...
Страница 46: ...3 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 55: ...Revision A Dash 3000 4000 Patient Monitor 4 1 2000966 339 4 Maintenance ...
Страница 56: ...4 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 117: ...Revision A Dash 3000 4000 Patient Monitor 5 1 2000966 339 5 Troubleshooting ...
Страница 118: ...5 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 149: ...Revision A Dash 3000 4000 Patient Monitor 6 1 2000966 339 6 Configuration ...
Страница 150: ...6 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 173: ...Revision A Dash 3000 4000 Patient Monitor 7 1 2000966 339 7 Calibration ...
Страница 174: ...7 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 182: ...8 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 226: ...9 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 236: ...9 12 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Theory Of Operation Block Diagram ...
Страница 278: ...9 54 Dash 3000 4000 Patient Monitor Revision A 2000966 339 Assembly Drawings Port Connections For your notes ...
Страница 279: ...Revision A Dash 3000 4000 Patient Monitor A 1 2000966 339 A Appendix A Electromagnetic Compatibility EMC ...
Страница 280: ...A 2 Dash 3000 4000 Patient Monitor Revision A 2000966 339 For your notes ...
Страница 291: ......