Suspend all monitoring and alarms for SpO

2

.

Choose whether or not the plethysmographic 
waveform appear on the screen.

Adjust or auto-set alarm limits. Select hi 
to change the high alarm limit and lo to 
change the low alarm limit. The system 
maintains appropriate separation between 
the high and low limits.

NOTE: The 

auto-set

 option is not available for SpO

2

. You cannot 

choose the 

auto-set

 option to set the alarm limits for oxygen 

saturation. However, when SpO

2

 is the source for HR/Pulse, using 

auto-set

 in the 

Setup HR/Pulse

 menu will set the alarm limits 

around the SpO

2

 PR value.

Heart Rate

Heart rate/pulse (HR/Pulse) in the Dash 2500 Patient  
Monitor can be derived from one of several sources. In 
descending order of priority, they are ECG, pulse oximetry 
(SpO

2

), and non-invasive blood pressure (NIBP). Instead of 

choosing a single source of HR/Pulse, you can activate the 
auto feature to choose the best available source.

•  Prepare patient

•  Connect monitoring lines

•  Select 

HR/Pulse

 from the 

setup parameters

 option in 

the 

Main Menu

•  Set up monitoring parameters

Choose the source of the HR/Pulse data to 
be displayed on the screen.
Adjust the volume of the tone that sounds 
each time a QRS complex is detected.

Adjust or auto-set alarm limits. Select hi 
to change the high alarm limit and lo to 
change the low alarm limit. The system 
maintains appropriate separation between 
the high and low limits. Choose auto-set to 
use the patient’s current readings in setting 
the high and low limits.

•  Manual mode: push the 

NIBP Go/Stop

 hardkey once to 

obtain immediate determination. 

•  Select 

NIBP

 from the 

setup parameters

 option in 

Main 

Menu

 to set or change limits or to set up auto mode. 

•  Auto mode: quick-push the 

NIBP Auto

 hardkey to begin 

auto BP determinations.  

 

•  Stat mode: push-and-hold the 

NIBP Auto

 hardkey to 

obtain continuous determinations for 5 minutes.

Suspend all monitoring and alarms for NIBP.

Configure a custom auto mode to meet the 
needs of your clinical situation.

Set time-cycled (auto mode) NIBP  
determinations.

Adjust or auto-set alarm limits. Choose hi to 
change the high alarm limit and lo to  
change the low alarm limit. You can set 
high and low limits individually for systolic, 
diastolic, and mean values. The system 
maintains appropriate separation between 
the high and low limits. Choose auto-set to 
use the patient’s current readings in setting 
the high and low limits.

 

 

 

 

  

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  

 

SpO

2

The SpO

2

 parameter in the Dash 2500 Patient Monitor is 

available in two different leading technologies: Nellcor

®

 and 

MASIMO SET

®

.

•  Select sensor

•  Plug sensor connector into SpO

2

 port

•  Apply sensor

•  Select 

SpO

2

 from the 

setup parameters

 option in the 

Main Menu

 

•  Set up monitoring parameters

NOTE: Attach the SpO

2

 sensor to the limb opposite the one 

with the blood pressure cuff.

NOTE: Remove nail polish and artificial nails.