February 2009
1-30
Programming Site Data
Entering User Functions
User functions enable you to program mathematical equations on
each measurement. You can also use any parameter in the meter to
calculate a different parameter. To enter the User Functions option,
scroll to the User Functions entry on the Program Menu and press
[ENTER]
. The screen appears similar to Figure 1-29 below. To step
through each parameter, press the [
T
] key. Refer to the
Program
Menu Map
, Figure 1-34 on page 1-37.
Figure 1-29: User Function Window
1.
The first prompt requests the function number, 1 through 8.
a.
Press
[ENTER]
to open the drop-down menu.
b.
Scroll to the desired function number (User1, User2, etc.).
c.
Press
[ENTER]
to confirm the entry.
2.
The next prompt asks you to create a label for the function.
Note:
The label corresponds to the measurement type (i.e., velocity
or temperature), the units symbol corresponds to the
measurement units (i.e., feet/sec or degrees F), and the Dec.
option elects the number of decimal places.
Done
Cancel
0
Function
Label
Units Sym
Dec.
Delete
Set User Function