CRxFlex Operating Manual V1.1
37
Environmental Conditions
Operation
Temperature recommended:
20 °C - 25 °C
Temperature allowed
15 °C - 30 °C
Maximum temperature change
0.5 °C/min.
Relative humidity
recommended
30 % - 60 %
Relative humidity allowed
15 % - 75 % (non condensed
Magnetic field
compliant with EN 61000-4-8,
Level 2
Sunlight exposure
Not to be operated in direct
sunlight, max. 2000 lux
Transport
IEC60721-3-2 (1997: class 2K2
and 2M2, with following
restrictions
Temperature
-25 °C to +55 °C
Vibration
5
– 200 Hz (vertical, longitudinal,
transversal axis
Electrical Data
Operating voltage
Auto-ranging power supply from
100-120 / 220-240 V
Operating current
2.0 / 1.0 A
Mains frequency
50 / 60 Hz
Mains fuse protection
Europe: min. 10 A, max. 16 A
USA & Japan: min. 10 A, max.
15A
Power consumption
Standby
max. 60W
During operation
max. 220W
Dimensions in cm [inch]
78.6 [30.9]
69.3 [27.2]
52.5 [20.7]
Dimensions in cm [inch]
69.3 [27.2]
52.5 [20.7]
100.2 [39.5]