![GE Corometrics 170 Series Скачать руководство пользователя страница 118](http://html.mh-extra.com/html/ge/corometrics-170-series/corometrics-170-series_service-manual_128350118.webp)
6-16
170 Series Monitor
Revision C
2000947-004
Functional Checkout Procedure:
Fetal Movement Detection Test
Fetal Movement Detection Test
This portion of the functional checkout procedure ensures the integrity of the fetal
movement detection circuitry and the heart rate channel of the recorder. (Refer to
.)
NOTE:
Fetal movement detection is an option which must be installed and enabled
for use on your monitor.
1. Connect the simulator’s US sub-cable to the
US
receptacle on the monitor.
2. Set the switches on the Model 325 Input Simulator according to
3. If not already on, press the monitor’s
Record
pushbutton.
4. Turn the simulator’s
Manual Adjustment
knob to input an ultrasound signal of
approximately 120 BPM. Verify the following on the monitor:
The FHR1 value is 120 BPM.
The FHR1 heartbeat indicator (
) flashes at a rate of 120 times per
minute.
Ultrasound audio volume from the rear panel speaker can be increased or
decreased using the left pair of
Volume
pushbuttons. (Set the volume to
the desired level.)
The recorder prints a continuous line at 120 BPM on the top grid of the
strip chart paper.
Fetal movement markers
are shown on for a duration of one second,
then off for eight seconds, then on for one second, etc.
The recorder prints the messages
US
and
FMD
on the center margin
of the strip chart paper after approximately 20 seconds.
Table 6-4. Fetal Movement Detection Test Simulator Settings
Section
Switch
Setting
US/FMD
Mode
US/FMD
Signal Level
MED
Rate
MANUAL
General
Pattern Memory
OFF
UA
Main
CMR
Mode
TOCO
Содержание Corometrics 170 Series
Страница 1: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 2: ......
Страница 3: ...Corometrics 170 Series SERVICE MANUAL MANUAL P N 2000947 004 REV C ...
Страница 18: ...1 10 170 Series Monitor Revision C 2000947 004 Safety Equipment Symbols For your notes ...
Страница 24: ...2 6 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 40: ...3 16 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 59: ...Revision C 170 Series Monitor 4 19 2000947 004 Setup Procedures Quick Reference Card ...
Страница 112: ...6 10 170 Series Monitor Revision C 2000947 004 Functional Checkout Procedure FECG Test Figure 6 1 FECG Ramp ...
Страница 117: ...Revision C 170 Series Monitor 6 15 2000947 004 Functional Checkout Procedure Ultrasound Test Figure 6 3 Ultrasound Ramp ...
Страница 182: ...7 50 170 Series Monitor Revision C 2000947 004 Serviceable Assemblies Boot ROM Error Codes For your notes ...
Страница 202: ...8 20 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 224: ...9 22 170 Series Monitor Revision C 2000947 004 For your notes ...
Страница 247: ......