Configuration
2026419-033E
CIC Pro™
6-45
5.
After making your selections, choose one of the following:
Click
OK
to apply your changes and close the
CIC Setup
window.
Click
Apply
to apply your changes without closing the
CIC Setup
window.
6.
Verify the new settings prior to monitoring patients.
Set the Telemetry Alarm Control Defaults
This option sets the telemetry default alarm limits and alarm level settings. In user
mode, all of the controls on the
Telemetry Alarm Control Defaults
tab are view-only.
You must be in the Service mode to set the
Telemetry Unit Defaults
at the CIC Pro
center.
Option Function
Default Location for this CIC
Set the print location for telemetry bed patient data:
Manual
: Designate the default manual graph location for telemetry patients.
Alarm
: Designate the default alarm graph location for telemetry patients.
Print WIndow
: Designate the default print window location for telemetry patients.
NOTE
These default locations are only used for telemetry beds and determine where patient data
prints for either manual or alarm conditions. Since a telemetry patient is not linked to a patient
monitor, these defaults are necessary to specify the destination for alarm and manual graph
printouts.
Waveforms
Designate the primary ECG lead for printing and enable or disable printing from subsequent ECG
leads.
ECG 1
: Designate the primary ECG lead for printing.
Waveform 2
to
Waveform 4
: Choose other ECG leads to print or choose
OFF
to disable printing
an ECG lead.
Transmitter Graph
Turn transmitter graph printing on or off.
Alarm Graph
Turn alarm graph printing on or off.
Event Marker Graph
Turn event marker graph printing on or off.
Display Lead
Set the primary ECG lead for display in the patient’s waveform window.
Arrhythmia
Enable or disable an arrhythmia analysis program.
Lead Analysis
Designate
Single-Lead
or
Multi-Lead
analysis for ECG and arrhythmia analysis.
ST Analysis
Enable or disable ST analysis.
Va Lead
Vb Lead
Va Lead/Vb Lead: Set the default for the V leads that will be monitored in these positions. A 6-lead
cable is required for multiple V-lead monitoring.
Detect Pace
Enable or disable pacer detection.
Patient Age
Set patient age.
Transmitter Alarm Pause
Turn transmitter alarm pausing on or off.
Alarm Pause Breakthrough
Turn transmitter alarm pause breakthrough on or off.
Event Marker
Turn event marker alert on or off.
Содержание CIC Pro
Страница 12: ...x CIC Pro 2026419 033E ...
Страница 13: ...2026419 033E CIC Pro 1 1 1 Introduction ...
Страница 27: ...2026419 033E CIC Pro 2 1 2 Equipment overview ...
Страница 53: ...2026419 033E CIC Pro 3 1 3 Licensing ...
Страница 60: ...3 8 CIC Pro 2026419 033E Licensing ...
Страница 61: ...2026419 033E CIC Pro 4 1 4 Service interfaces ...
Страница 74: ...4 14 CIC Pro 2026419 033E Service interfaces ...
Страница 75: ...2026419 033E CIC Pro 5 1 5 Installation ...
Страница 94: ...5 20 CIC Pro 2026419 033E Installation ...
Страница 95: ...2026419 033E CIC Pro 6 1 6 Configuration ...
Страница 187: ...2026419 033E CIC Pro 7 1 7 Checkout procedures ...
Страница 208: ...7 22 CIC Pro 2026419 033E Checkout procedures ...
Страница 209: ...2026419 033E CIC Pro 8 1 8 Troubleshooting ...
Страница 241: ...2026419 033E CIC Pro 9 1 9 Field replaceable units FRUs ...
Страница 243: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 3 Hard drive Flash drive ...
Страница 244: ...9 4 CIC Pro 2026419 033E Field replaceable units FRUs Rack mount ...
Страница 245: ...Field replaceable units FRUs 2026419 033E CIC Pro 9 5 Mounting options ...
Страница 265: ...2026419 033E CIC Pro 10 1 10 Preventive maintenance ...
Страница 279: ...2026419 033E CIC Pro 11 1 11 Reload software ...
Страница 288: ...11 10 CIC Pro 2026419 033E Reload software ...
Страница 289: ...2026419 033E CIC Pro 12 1 12 Upgrade software ...
Страница 298: ...12 10 CIC Pro 2026419 033E Upgrade software ...
Страница 299: ...2026419 033E CIC Pro A 1 A Electromagnetic compatibility ...
Страница 305: ...2026419 033E CIC Pro B 1 B Backup and restore backed up data modules ...
Страница 309: ...Device compatibility 2026419 033E CIC Pro C 1 C Device compatibility ...
Страница 312: ...C 4 CIC Pro 2026419 033E Device compatibility ...
Страница 313: ......