Flow control test
The Link-25 system is tested during the "
" part of the
checkout procedure. This procedure is an alternate manual test for
significant malfunction of the Link-25 system.
This procedure tests for significant malfunction of the
Link-25 system. It does not confirm the proper calibration
of the Link-25 system. Perform periodic calibrations using
an accurate and properly calibrated O2 monitor as
recommended in the "
" section.
1.
Connect the pipeline supplies or slowly open the cylinder valves.
2.
Turn all flow controls fully clockwise for minimal flow.
3.
(ACGO option only.) Set the ACGO switch to circle circuit
position.
4.
(Aux O2+Air only.) Set the Aux O2+Air switch to circle circuit
position.
5.
Set the system to On.
Do not use the system if the Low battery or any ventilator failures
occur.
Keep the Link-25 system engaged. Adjust only the test
control for the following steps.
•
Test N2O first and then O2.
•
The O2 cell must be correctly calibrated.
6.
Test the Link-25 system by increasing the N2O flow.
•
Slowly turn the N2O flow control counterclockwise.
•
Increase the N2O flow as specified in the following table
and make sure the O2 flow is as specified.
N2O flow l/min
O2 flow greater than l/min
0.8
0.2
2
0.5
4
1.0
10
2.5
7.
Test the Link-25 system with O2 flow decreasing.
•
Set the N2O flow to 9 l/min.
Note
WARNING
Note
WARNING
9 User maintenance
2076152-001
9-9
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...