Minimum Frequency
Maintenance
As necessary
•
Install new cylinder gaskets on cylinder yokes.
•
Replace the absorbent in the canister.
•
Empty and clean the overflow trap on the optional suction regulator.
•
Replace the circuit O2 cell. (Under typical use the cell meets
specifications for 2 years. )
•
Replace the autoclavable flow sensors. (Under typical use the sensor
meets specifications for a minimum of 1 year.)
•
Calibrate the airway modules every 6 months or when there are
indications of errors in the gas readings. Calibrate airway modules that
get extensive usage every 2 months.
•
Inspect and clean the fan filters (display unit, rear side of the machine
and airway module).
•
Empty D-fend Pro water trap on the airway module.
•
Drain the vaporizers and discard the agent as instructed in the
vaporizer User's Reference Manual (This is not necessary for the
Tec
™
6 series vaporizers.).
Authorized service personnel
This is the minimum level of maintenance recommended. Local
regulations may contain additional maintenance requirements.
Comply with all local regulations which meet or exceed this minimum
level of maintenance.
.
Minimum Frequency
Maintenance
12 months
Have an authorized service representative complete the scheduled service
maintenance checks, tests, calibrations, and parts replacements as defined
in the Technical Reference manual.
9 User maintenance
2076152-001
9-5
Содержание Carestation 620
Страница 1: ...Carestation 620 650 650c A1 User s Reference Manual Software Revision 01...
Страница 12: ...Carestation 620 650 650c A1 x 2076152 001...
Страница 57: ...3 Push the ComWheel or select the quick key to confirm the change 2 System controls and menus 2076152 001 2 23...
Страница 58: ...2 24 2076152 001...
Страница 98: ...3 40 2076152 001...
Страница 102: ...4 4 2076152 001...
Страница 128: ...6 12 2076152 001...
Страница 150: ...7 22 2076152 001...
Страница 178: ...4 Tighten all thumb screws on the rear of the mounting plate AC 24 265 Carestation 620 650 650c A1 8 28 2076152 001...
Страница 182: ...8 32 2076152 001...
Страница 190: ...12 Calibrate the new O2 cell according to the Circuit O2 cell procedure Carestation 620 650 650c A1 9 8 2076152 001...
Страница 196: ...9 14 2076152 001...
Страница 246: ...11 44 2076152 001...
Страница 250: ...3 Set the cost for each agent 4 Select Back to return to the Gas Usage menu Carestation 620 650 650c A1 12 4 2076152 001...
Страница 268: ...Carestation 620 650 650c A1 User s Reference Manual English 2076152 001 2021 05 Rev S...