Neuromuscular transmission
2.
Check that the electrodes are correctly positioned on the ulnar nerve and the
message
Supramax search
is displayed. Ensure that you get a stimulus response.
If the supramaximal stimulus current is not found, the message
Supramax not
found
is displayed.
NMT measurement setup
NMT equipment to patient connection
1.
Module with NMT measurement capability
2.
NMT sensor cable
3.
MechanoSensor or Pediatric MechanoSensor lead wire set
4.
Electrode, white lead connection site for nerve stimulation
5.
Electrode, brown lead connection site for nerve stimulation
6.
ElectroSensor leadwire set
7.
White stimulating electrode
8.
Brown stimulating electrode
9.
Electrode, black lead connection site, ground
10. Electrode, green lead connection site, recording muscle-contraction effect
11. Electrode, red lead connection site, recording muscle-contraction effect
NMT module keys
There are two keys on the module:
Start-up
●
Starts the search for supramaximal current and reference level.
●
Proceeds with the selected measurement cycle.
Stop Continue
●
Interrupts monitoring.
●
Restarts monitoring of the same patient.
Preparing the patient for NMT measurement
1.
Connect the NMT sensor cable to the module.
2.
Clean the skin at the NMT application area.
3.
When applying the electrodes, make sure that the entire electrode surfaces make
an optimal contact to the skin.
4.
Always connect the white NMT stimulation lead to the proximal electrode.
2062971-001
CARESCAPE Modular Monitors
363
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...