Airway gases
3.
Adapter with sampling line connector
4.
Sampling line connector on the water trap
Airway gases equipment to patient connections with
E-miniC, critical care setup
1.
E-miniC module
2.
Gas sampling line
3.
Adapter with sampling line connector
4.
Sampling line connector on the water trap
Airway gases measurement setup
1.
Make sure that the water trap container is empty and properly attached.
2.
Connect the gas sampling line to the sampling line connector on the water trap.
3.
Connect the sample gas outlet to gas scavenging if N
2
O or volatile agents are
used.
4.
Turn on the monitor or connect the module to the monitor. The monitor performs
a self-check for the module when the module is connected. Automatic agent
identification is activated in those modules that have this feature.
5.
Wait until the message
Calibrating
disappears.
6.
Connect the sampling line to the airway adapter or the airway adapter to the
ventilator circuit. Position the adapter with the sampling port upwards to minimize
the amount of condensed water possibly entering the sampling line.
7.
Check that the airway adapter connections are tight and that the adapter is
operating properly.
NOTE
Check that the sample line is connected to the water trap
before connecting the module to the monitor or turning on
the monitor.
NOTE
To minimize the amount of dust drawn into the gas sampling
system, always keep the water trap connected to the module.
When gas measurement is not in use, you can disconnect the
module from the monitor to eliminate the operating sound
of the gas pump.
286
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...