Pulse oximetry
system is functional, they are incapable of providing the data required to properly
evaluate the accuracy of the system’s SpO
2
measurements.
Many functional testers and patient simulators have been designed to interface with
the pulse oximeter’s expected calibration curves, and may be suitable for use with
the pulse oximeter monitor and sensors. Not all devices, however, are adapted for
use with all digital calibration systems. While this will not affect use of the simulator
for verifying system functionality, displayed SpO
2
measurement values may differ
from the setting of the device. For a properly functioning monitor, this difference
will be reproducible over time and from monitor to monitor within the performance
specifications of the test device.
If a particular calibration curve is accurate for the combination of a pulse oximeter
monitor and sensor, a functional test device can measure the contribution of a
monitor to the total error of a monitor/sensor system. The functional test device can
then measure how accurately a particular pulse oximeter monitor is reproducing
that calibration curve.
Using the SpO₂ measurement
Primary and secondary SpO
2
measurement sources
It is possible to measure SpO
2
from two different measurement sources
simultaneously. The primary SpO
2
source is labeled
SpO2
and the secondary SpO
2
source is labeled
SpO2(2)
.
The following table shows the acquisition modules that may be used as primary and
secondary SpO
2
measurement sources.
Primary SpO
2
source
Compatible secondary SpO
2
source(s)
PDM, TRAM, PSM
●
Tram-Rac SpO
2
module
●
E-NSATX
●
E-MASIMO
●
Unity Network ID connectivity device
E-NSATX, E-MASIMO
●
Unity Network ID connectivity device
Unity Network ID connectivity
device
●
Tram-Rac SpO
2
module
●
E-NSATX
●
E-MASIMO
E-NSATX and E-MASIMO modules require a PDM with no
SpO
2
sensor connected in order to work as a secondary
SpO
2
source when the Unity Network ID connectivity device
is the primary SpO
2
source.
Changing the SpO
2
waveform size
NOTE
All other modules except PSM.
1.
Select the SpO
2
parameter window.
2.
Select the
SpO2
or
SpO2(2)
tab.
3.
Choose the size from the
Size
list:
1x
,
2x
,
4x
, or
8x
.
208
CARESCAPE Modular Monitors
2062971-001
Содержание CARESCAPE
Страница 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Страница 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Страница 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Страница 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Страница 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Страница 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Страница 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Страница 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Страница 322: ...CO2 with CAPNOSTAT Mainstream CapnoFlex LF and Dual CO2 modules 322 CARESCAPE Modular Monitors 2062971 001 ...
Страница 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Страница 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Страница 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Страница 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Страница 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Страница 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Страница 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Страница 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Страница 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Страница 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Страница 547: ...content ...