Troubleshooting
145
2106778-001
11.5.3
Incorrect system time
11.5.4 License issues
Problem
Possible cause
Recommended action
System time is incorrect when
patient monitor is not
connected to network.
CPU timekeeper battery empty
Replace CPU timekeeper battery.
Time not configured properly
Configure date and time, see
System time is incorrect when
patient monitor is connected to
MC Network.
Network device time
synchronization error.
When adding a new device to the
CARESCAPE Network, the existing devices
on the CARESCAPE Network will
synchronize to the new device’s time. To
prevent potential time synchronization
issues, you should set the new device’s
time to be as close as possible to the time
(within one minute or less) used by the
existing GE devices on the CARESCAPE
Network.
System time is incorrect when
patient monitor is connected to
S/5 network.
iCentral time incorrect.
Configure iCentral time master.
Problem
Possible cause
Recommended action
Unable to perform a function or
a feature is not available.
- A license has not been
purchased for the feature.
- The trial license has expired
for the feature.
- The license is not installed
properly.
See License management chapter.
Содержание CARESCAPE Monitor B850
Страница 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Страница 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Страница 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Страница 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Страница 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Страница 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Страница 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Страница 157: ...Troubleshooting 149 2106778 001 Problem No waveforms or parameters are displayed at the CARESCAPE Cental Station ...
Страница 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Страница 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Страница 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Страница 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Страница 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Страница 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Страница 191: ......