Appendix A – Maintenance and Troubleshooting
2002938-002E
Cardiolab® II Plus Amplifier Manual
A-5
Clean the FOC contacts with a non-abrasive, lint-free cloth and
isopropyl alcohol.
Replace the fiber optic cable.
If the waveforms only exist on one channel, try replacing either the
catheter, catheter extender, or catheter junction box. You may also
try another Catheter Input Module.
Noise on Intracardiac Signals
The noise should resemble 60 Hz or some other relatively consistent
interference which are not associated with stim artifact.
Go to the amplifier test selection under Administration and perform
a hardware reset on the amplifier (this is equivalent to turning the
amplifier off and on, which would be the second option).
Verify that the catheter is correctly plugged into the Catheter Input
Module (CIM) and that these poles have been correctly entered into
the study configuration.
Remove/replace the catheter extender cables, especially if they have
been re-sterilized.
For real-time testing during a case:
Make duplicate channels
If the noisy signal is set to poles x and y, set up another channel (one
with a high quality intracardiac signal, (for example, HRA or RVA)
with the same poles and see if the noise is present there as well. This
tests the channel.
While keeping the channels consistent, try inputting the catheters to
different poles on the CIM and see if the noise follows the signal. This
tests the CIM poles.
NOTE
If the last two tests make no difference, the problem may exist with
the catheter itself. Try replacing the catheter.
Detailed Testing: Amplifier Channel Test
Input a simulated signal into poles 1 and 2 of the CIM block A and
connect the ground of the simulator to the RL limb lead
In the study configuration set all intracardiac channels to:
Poles 1 and 2
Содержание Cardiolab II Plus
Страница 1: ...Service Manual 2002938 002 Revision E Cardiolab II Plus Amplifier Manual ...
Страница 8: ...i iv Cardiolab II Plus Amplifier Manual 2002938 002E ...
Страница 9: ...2002938 002E Cardiolab II Plus Amplifier Manual 1 1 1 Introduction ...
Страница 26: ...1 18 Cardiolab II Plus Amplifier Manual 2002938 002E Introduction ...
Страница 27: ...2002938 002E Cardiolab II Plus Amplifier Manual 2 1 2 Equipment Overview ...
Страница 36: ...2 10 Cardiolab II Plus Amplifier Manual 2002938 002E Equipment Overview ...
Страница 37: ...Equipment Overview 2002938 002E Cardiolab II Plus Amplifier Manual 2 11 ...
Страница 45: ...Equipment Overview 2002938 002E Cardiolab II Plus Amplifier Manual 2 19 Stimulator Input Cable ECG 10 Lead Set ...
Страница 46: ...2 20 Cardiolab II Plus Amplifier Manual 2002938 002E Equipment Overview 10 Lead ECG Cable Power Cord ...
Страница 47: ...2002938 002E Cardiolab II Plus Amplifier Manual 3 1 3 Operation ...
Страница 51: ...2002938 002E Cardiolab II Plus Amplifier Manual A 1 A Appendix A Maintenance and Troubleshooting ...
Страница 65: ...2002938 002E Cardiolab II Plus Amplifier Manual B 1 B Appendix B Technical Information ...
Страница 66: ...B 2 Cardiolab II Plus Amplifier Manual 2002938 002E Appendix B Technical Information Technical Specifications ...
Страница 67: ...Appendix B Technical Information 2002938 002E Cardiolab II Plus Amplifier Manual B 3 ...
Страница 68: ...B 4 Cardiolab II Plus Amplifier Manual 2002938 002E Appendix B Technical Information ...
Страница 69: ......