Chapter 4 - Room Planning
Page 93
nning
Figure 4-23 Typical Scatter Survey (Body Filter)
0.65
0.65
1.3
1.3
2.6
2.6
5.2
5.2
Technique:
140 kV
100mA
1 Sec
4 x 5.00mm
50cm
50cm
BODY SCATTER PHANTOM
0.65
0.65
1.3
1.3
2.6
2.6
5.2
5.2
Technique:
140 kV
100mA
1 Sec
4 x 5.00mm
50cm
50cm
BODY SCATTER PHANTOM
ISO–Contour Levels: 0.65, 1.3, 2.6 and 5.2 µGy/Scan
ISO–Contour Levels: 0.65, 1.3, 2.6 and 5.2 µGy/Scan