Chapter 4 - Room Planning
Page 83
nning
Section 7.0
Minimum Dimensions and Clearances
7.1
Options
Ceiling Pedestal mount lowest point to floor (Injector or Monitor) 2134.0 mm (84.0 in)
7.2
System Operation
7.3
Injector Control
A suitable work area, which is within reach of the operator's console, should be provided for
placement of the injector control.
Wall mounted, ceiling mounted and pedestal units need cables to be routed from the gantry area to
the console area. Injectors AC power is supplied with Accessory I/F hardware kit.
Note:
Injector cables should not be routed with the system cables.
Mounts are available in different configurations and lengths. Refer to Injector documentation for
detailed installation instructions.
SYSTEM OPERATION
MM
INCHES
Finished ceiling to floor (recommended)
2743.0mm
108.0 in.
Finished ceiling to floor (minimum)
2286.0mm
90.0 in.
HPower Table max extension head end w/extender from Center Line
2030.0mm
80.0 in.
HPower Table extension head end w/extender to obstruction
152.0mm
6.0 in.
HPower Table in lowest pos. w/cradle @home pos. to Center Line
3209.0mm
126.5 in.
GT1700 Table max extension head end w/extender from Center Line
1712.0 mm
67.0 in.
GT1700 Table extension head end w/extender to obstruction
150.0 mm
6.0 in.
GT1700 Table in lowest pos. w/cradle @home pos. to Center Line
3209.0 mm
126.5 in.
GT1700 Table in lowest pos. w/cradle @home pos. to surface of Gantry front cover
2744.0 mm
108.0 in.
Back of GOC4/LCGOC/AIO Console to wall
152.0mm
6.0 in.
Back of TIO/NIO16 Console to wall
96.0mm
4.0 in.
Back of PDU to wall
152.0mm
6.0 in.