Page 102
Section 5.0 - System Component Noise Levels
4.8
Equipment EMI “Envelopes”
Figure 5-1 Sample Room Layout, showing approximate EMI requirements
Section 5.0 System Component Noise Levels
Maximum Gantry Audible Noise Level
The maximum ambient noise levels is produced by the
gantry during a CT scan acquisition. It is less than 75 dBA when measured at a distance of one
meter from the nearest gantry surface, in any direction.
Patient
Table
Gantry
Power
Distribution
Unit
Operator's
Console
Laser Camer
(option)
GE Storage
Cabinet
(option)
Injector
Control
(option)
Slave
Monitor
(option)
TV Monitor
(not GE)
Counter w/sink and
Base Cabinets
TV
Camera
(not GE)
Note: Drawing is not to scale
UPS
Range of PDU's
EMI is
0.3m (12")
or
1m (39")
in ALL directions.
Do not place
any electronics
within this area!
(Not GE Supplied)
1m (39")
TV Monitor
(not GE)
Operator's
Console