GE H
EALTHCARE
D
IRECTION
5141177-100, R
EVISION
14
B
RIGHT
S
PEED
E
LITE
, E
DGE
, E
XCEL
: P
RE
-I
NSTALLATION
Chapter 7 - Delivery and Storage Requirements
Page 123
e
ry
Da
ta
Figure 7-3 Package Symbols (Storage)
Section 9.0
Extreme Temperature Transportation and Deliveries
NOTICE
Failure to adhere to Extreme Temperature Transportation and Delivery requirements will
likely result in equipment damage.
Extreme temperatures should be avoided during system transportation and delivery.
Extreme temperatures consist of temperatures below -18º C (0º F), or above 49º C (120º F), without
humidity control.
When transporting the CT system, prevent extended exposure of the system to temperatures or
humidity outside of the following specifications:
•
Temperature: -40º to +70º C (-40º to +158º F)
•
Humidity: 5% to 95%
NOTICE
Component Freezing occurs if CT system is exposed to temperatures below -18º C (0º F) for
a period longer than two days.
Allow a minimum of 12 hours for the CT system to adjust to ambient room temperature, prior
to installation.
4
27
4
20-60%
700