Safety labels
The table below describes the various safety labels that may be found on BioProcess
LPLC or MPLC columns or on associated BioProcess equipment.
Description
Symbol/text
Warning! Read the user documentation before using
the system. Do not open any covers or replace parts
unless specifically stated in the user documentation.
WARNING! Max operating pressure as stated on the
label.
EMERGENCY STOP label, yellow with black text.
(emergency stop button is red).
See Section 2.3 Emergency procedures, on page 29 for
further information regarding the emergency stop.
IMPORTANT! Before service/maintenance or return to
GE Healthcare, clean the equipment and accompany it
with a decontamination statement, specifying
substances with which it has been in contact during
use and the method of cleaning.
CAUTION! Do not use any solvent solution when
cleaning transparent plastic items.
CAUTION! Label stating pressure control valve settings
for instrument air supply.
CAUTION!
Pressure control valve
shall be set to
5.5 – 7 bar g
for instrument air
supply
28
LPLC and MPLC Columns Operating Instructions 28-9638-66 AC
2 Safety instructions
2.2 Labels