Chapter 6
Troubleshooting
Alarm/Troubleshooting Guide
Symptom
Likely causes
Troubleshooting steps
Unit Overheated
Indicator light is
on.
Clogged air filter,
blocked vents,
fan failure.
1. Turn off the standby switch and
disconnect the power cord from the
outlet.
Note
: It can take up to 10 minutes for the
thermal cutout switch to reset and allow the
unit to be used again.
2. Make sure that the air filter in the back of
the unit and the air vents on either side
of the unit are unobstructed. If the air
filter needs to be removed, take the unit
out of service and seek assistance from
service personnel.
3. Plug the power cord into an outlet. Make
sure a fiberoptic light pad is attached.
Turn the standby switch on.
4. Verify that the fan is on and allow the
unit to run with the fiberoptic light pad
attached.
5. If the fan is not running while the unit is
turned on or if the Unit Overheated
indicator light turns on again, remove the
unit from service and seek assistance
from service personnel.
LED failure
indicator light is
blinking.
At least one of
the three LED
pairs is not
operating.
1. LED module needs to be replaced by
service personnel.
Note:
The system may still be used until the
LED module is replaced. However the
intensity will be lower than the specification.
M1110147 Rev.
5
© 2007 Datex-Ohmeda. All rights reserved.
6-1
Содержание BiliSoft
Страница 2: ...ii M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 4: ...iv M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 6: ...vi M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 8: ...viii M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 10: ...1 2 M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 18: ...3 4 M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 30: ......
Страница 34: ...6 4 M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 40: ...7 6 M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 42: ...8 2 M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...
Страница 46: ...10 2 M1110147 Rev 5 2007 Datex Ohmeda All rights reserved This page intentionally left blank ...