7-1
Chapter 7
Specifications
Note: All specifications are nominal and are subject to change without notice.
Electrical Specifications
Device Insulation Class
Class II Double Insulated
Applied Parts (Light Pad)
Type BF
Input Voltage
100 – 240 V
~
, 50/60 Hz
Input Current
0.25A – 0.75A
Leakage Touch Current
< 100 μA in normal condition
< 500 μA in single fault condition
Environmental Operating Conditions
Temperature
+5°C to +35°C (41°F to 95°F).
Humidity
15% to 90% RH, non-condensing, but not requiring a water
vapor partial pressure > 5 kPa
Atmospheric pressure
70 kPa to 106 kPa (10,000 ft. to -1,250 ft.)
Light Box
IP21
Light Pad
IPX4
Transportation Conditions & Storage Conditions (while sealed in shipping carton)
Temperature
-25°C to +5°C (-13°F to 41°F) without humidity control.
Temperature
+5°C to +35°C (41°F to 95°F) up to 90% RH,
non-condensing
Temperature
+35°C to +70°C (95°F to 158°F) at a water vapor
pressure up to 5 kPa
Содержание BiliSoft 2.0
Страница 1: ...BiliSoft 2 0 Phototherapy System Lighting by LumitexTM Operation Manual ...
Страница 8: ...viii This page intentionally left blank ...
Страница 12: ...xii This page intentionally left blank ...
Страница 14: ...xiv This page intentionally left blank ...
Страница 18: ...xviii This page intentionally left blank ...
Страница 24: ...1 6 This page intentionally left blank ...
Страница 32: ...2 8 This page intentionally left blank ...
Страница 36: ...3 4 This page intentionally left blank ...
Страница 43: ...5 1 Chapter 5 This chapter has been intentionally left blank ...
Страница 44: ...5 2 This page intentionally left blank ...
Страница 46: ...6 2 This page intentionally left blank ...
Страница 56: ...9 6 This page intentionally left blank ...
Страница 58: ...A1 2 This page intentionally left blank ...
Страница 71: ...A2 13 This page intentionally left blank ...