Vaporizer
The electronically controlled vaporizer consists of the internal
electronic control unit and the Aladin agent cassette. The agent
cassettes are color coded, have indexed filling ports, and are
magnetically coded for each agent. The electronic control unit
governs the flow through the agent cassette and the agent
concentration in the fresh gas flow.
Both the Aladin2 and the Aladin cassettes can be used on this
system.
Remove the cassette from the active bay when not administering
anesthetic agent. Store cassettes in the cassette storage bay when
they are not in use. Store the cassettes and agent at the same
temperature as the system.
Aladin2 cassette variants
All Aladin2 cassettes have electronic agent level sensing. The agent
level shows graphically in the agent settings area of the screen. If the
electronic agent level sensing is not functioning, the agent level
unknown symbol shows on the screen. In this case, refer to the liquid
level indicator.
Some Aladin2 cassettes have internal temperature sensing. An
enhanced temperature sensing symbol shows on the front of the
cassette and the symbol shows in the agent settings area of the
screen for cassettes that have internal temperature sensing.
There are three types of Aladin2 cassette filler systems. Enflurane
and isoflurane use a color-coded, Easy-Fil mechanism. Sevoflurane
cassettes are available with a color-coded, Easy-Fil or Quick-Fil
mechanism. The desflurane cassettes have a filling mechanism that
is compatible with Saf-T-Fil desflurane bottles.
Sevoflurane Quik-Fil Aladin2 cassettes are not available in all
countries.
Anesthetic agent
Filling system
Color code
Enflurane
Easy-Fil
Orange
Isoflurane
Easy-Fil
Purple
Sevoflurane
Easy-Fil or Quik-Fill
Yellow
Desflurane
Compatible with Saf-T-Fill
Blue
Note
Aisys CS²
13-2
2067226-001
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...