Display optional output
ports
Cables used for EMC testing
Monitor On/STBY
Standard 9-pin D-subminiature male
connector with shielded cable - 1.8 m
maximum length.
Guidance and manufacturer’s declaration - electromagnetic
emissions
The system is suitable for use in the specified electromagnetic
environment. The customer and/or user of the system should assure
that it is used in an electromagnetic environment as described below.
.
Emissions test
Compliance
Electromagnetic environment guidance
RF emissions
CISPR 11
Group 1
The system uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class A
The system is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic
emissions IEC
61000-3-2
Class A
The system is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Voltage
fluctuations/flicker
emissions IEC
61000-3-3
Complies
The system is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration - electromagnetic
immunity
The system is suitable for use in the specified electromagnetic
environment. The customer and/or user of the system should assure
that it is used in an electromagnetic environment as described below.
11 Specifications and theory of operation
2067226-001
11-47
Содержание auisys cs2
Страница 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Страница 25: ...Abbreviation Definition VCV Volume control ventilation Vol Volume 1 Introduction 2067226 001 1 13 ...
Страница 32: ...1 20 2067226 001 ...
Страница 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Страница 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Страница 90: ...3 42 2067226 001 ...
Страница 94: ...4 4 2067226 001 ...
Страница 132: ...7 20 2067226 001 ...
Страница 149: ...2 Auxiliary O2 flow control Figure 8 5 Auxiliary O2 flowmeter 8 Setup and connections 2067226 001 8 17 ...
Страница 164: ...9 8 2067226 001 ...
Страница 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Страница 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Страница 262: ...13 20 2067226 001 ...
Страница 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...