Viewing real-time patient data
2001989-200E
ApexPro™
7-21
Lethal arrhythmia conditions
When
Lethal
arrhythmia is selected, the following conditions (as defined for
Full
arrhythmia) are detected:
ASYSTOLE
VFIB/VTAC
V TACH
(defaults to the
Crisis
level, but can be moved to a different level)
BRADY
(if the
Patient Age
range selected is
0-2 years
or
3-10 years
)
AFIB identification
NOTE
While in combo mode, AFIB trending is only available when the bedside monitor
supports this feature. If your bedside monitor does not support AFIB trending,
this feature is not available. Contact your sales/service representative for more
information.
Atrial fibrillation (AFIB) is characterized by random, chaotic, low-amplitude
deflections of the supraventricular component of the ECG waveform, resulting in
irregular timing of QRS complexes and an absence of uniform P waves preceding the
QRS complex.
V TACH
Adult
—Ventricular tachycardia occurs when a run of six or more
ventricular beats is detected with an average heart rate greater than or
equal to 100 beats per minute.
0-2 years
—Occurs when a run of six or more ventricular beats is
detected with an average heart rate greater than or equal to 160 beats
per minute.
3-10 years
—Occurs when a run of six or more ventricular beats is
detected with an average heart rate greater than or equal to 140 beats
per minute.
11-13 years
—Occurs when a run of six or more ventricular beats is
detected with an average heart rate greater than or equal to 130 beats
per minute.
VT > 2
Adult
—Ventricular tachycardia >2 occurs when a run of ventricular
beats is detected with a duration of less than six beats but longer than
two beats and with an average heart rate that is greater than or equal
to 100 beats per minute.
0-2 years
—Occurs when a run of ventricular beats is detected with a
duration of less than six beats but longer than two beats and with an
average heart rate that is greater than or equal to 160 beats per
minute.
3-10 years
—Occurs when a run of ventricular beats is detected with a
duration of less than six beats but longer than two beats and with an
average heart rate that is greater than or equal to 140 beats per
minute.
11-13 years
—Occurs when a run of ventricular beats is detected with
a duration of less than six beats but longer than two beats and with an
average heart rate that is greater than or equal to 130 beats per
minute.
Содержание ApexPro
Страница 13: ...2001989 200E ApexPro 1 1 1 Introduction ...
Страница 31: ...2001989 200E ApexPro 2 1 2 Equipment overview ...
Страница 36: ...2 6 ApexPro 2001989 200E Equipment overview ...
Страница 37: ...2001989 200E ApexPro 3 1 3 Equipment setup ...
Страница 50: ...3 14 ApexPro 2001989 200E Equipment setup ...
Страница 51: ...2001989 200E ApexPro 4 1 4 System setup ...
Страница 69: ...2001989 200E ApexPro 5 1 5 Alarms ...
Страница 86: ...5 18 ApexPro 2001989 200E Alarms ...
Страница 87: ...2001989 200E ApexPro 6 1 6 Managing patients ...
Страница 109: ...2001989 200E ApexPro 7 1 7 Viewing real time patient data ...
Страница 144: ...7 36 ApexPro 2001989 200E Viewing real time patient data ...
Страница 145: ...2001989 200E ApexPro 8 1 8 Viewing stored patient data ...
Страница 166: ...8 22 ApexPro 2001989 200E Viewing stored patient data ...
Страница 167: ...2001989 200E ApexPro 9 1 9 Printing ...
Страница 175: ...2001989 200E ApexPro A 1 A Abbreviations and symbols ...
Страница 183: ...2001989 200E ApexPro B 1 B Customized defaults worksheet ...
Страница 186: ...B 4 ApexPro 2001989 200E Customized defaults worksheet ...
Страница 187: ...2001989 200E ApexPro C 1 C Maintenance ...
Страница 194: ...C 8 ApexPro 2001989 200E Maintenance ...
Страница 195: ...2001989 200E ApexPro D 1 D Troubleshooting ...
Страница 206: ...D 12 ApexPro 2001989 200E Troubleshooting ...
Страница 207: ...2001989 200E ApexPro E 1 E Technical specifications ...
Страница 214: ...E 8 ApexPro 2001989 200E Technical specifications ...
Страница 217: ......