background image

Section 2: Getting Started 

2023486-201 Rev C 

© 2007 General Electric Company 

Page 19 of 53

 

CONTROL BUTTONS 

The AED Pro has two buttons. 
 

SHOCK BUTTON 

 

The SHOCK button is located at the far right of the control panel and serves as an indicator to 
notify the user that the unit is ready to shock and as a button to deliver the shock. 
 
The word SHOCK and the shock button will illuminate RED when the AED Pro is ready to deliver a 
defibrillation shock to the patient.  

 
 

MANUAL OVERRIDE BUTTON 

 

The MANUAL OVERRIDE button is located at the far left of the control panel and converts the  
device from automated mode to manual. This feature should only be used by medical professionals 
trained in manual defibrillation. 

 
 
 
 

MANUAL OVERRIDE 

 

•  Lift the cover to access the button. 
 
•  Converts to manual standby mode when pushed once, a voice prompt “

Press Manual Button Again 

to Confirm

”, will be heard. Converts to manual mode when MANUAL button is pressed again.

 

 

•  If the rescuer does not confirm within 30 seconds of the capacitors charging, the AED will revert 

back to AED Mode. 

 
•  If the Medical Director has disabled this feature in MDLink, an icon indicating No MANUAL MODE 

will appear in the bottom left of the display 

 

SETTING THE AED PRO INTERNAL CLOCK 

The internal clock is preset at Central Standard Time and should be reset to the correct date and local time. The AED Pro 
will automatically adjust itself for daylight savings time. This feature can be turned off using the ServiceLink software. To 
set the clock, you will need a PC with Windows 95 or later operating system, RescueLink software installed, an IR port on 
the PC, and an IR adapter as specified below. 
 
To set the clock settings: 
 

 

Open the lid and remove pads from the pads socket. 

 

Ensure that the PC is set at the correct local time and date. 

 

Point IR port on the AED Pro to IR eye on the PC and select G3 Pro. 

 

Run the RescueLink software on the PC. 

 

Verify that the voice prompt states “Communications Mode”. 

 

Click Communications on the main menu. Select AED Pro Date and Time. 

 

Click on the Get button to review the current time in the AED Pro. 

 

If the time and date is incorrect, click Set to set new time and date. The AED Pro date and time will 
automatically be updated to the PC’s time and date. 

 

Reinstall pads per instructions on page 18. 

 

Close the lid. 

 
 

Note:

 The IR port on the AED Pro is designed to work with IR adapter ACT-IR220LN115 from ACTiSys 

Corp. on Windows based PCs only. Please contact customer service to order, P/N 162-0108-001. Other IR 
products may interfere with the transmission and are not for use with the AED Pro.

 

Содержание AED Pro

Страница 1: ...Operator s Manual 2023486 201rC...

Страница 2: ...AND CONDITIONS 4 SAFETY ALERT DESCRIPTIONS 4 SYMBOL DESCRIPTIONS 6 SAFETY AND PERFORMANCE STANDARDS 9 OPERATOR TRAINING REQUIREMENTS 10 SECTION 2 GETTING STARTED 11 OVERVIEW 11 UNPACKING AND INSPECTI...

Страница 3: ...F TESTS 35 INDICATOR TROUBLESHOOTING TABLE 36 SCHEDULED MAINTENANCE 36 AUTHORIZED REPAIR SERVICE 37 FREQUENTLY ASKED QUESTIONS 38 SECTION 6 TECHNICAL DATA 39 OVERVIEW 39 PARAMETERS 39 RHYTHMX AED ECG...

Страница 4: ...liver a shock by pushing the SHOCK button The AED Pro can also provide non diagnostic ECG monitoring INDICATIONS FOR USE INTENDED USE The AED PRO with STAR Biphasic Waveform is intended to be used by...

Страница 5: ...ion Hazard Do not use the AED Pro in the presence of flammable gases including concentrated oxygen to avoid possible explosion or fire hazard WARNING Shock Hazard Defibrillation shock current flowing...

Страница 6: ...not operate wireless radiotelephones within 2 meters of the AED Pro turn power OFF to the radiotelephone and other like equipment near the incident CAUTION Possible Interference with Implanted Pacema...

Страница 7: ...dards and compliances associated with the AED Pro and its use Dangerous Voltage The defibrillator output has high voltage and can present a shock hazard Please read and understand all safety alerts in...

Страница 8: ...he remainder of this manual Use pads by this date install battery by this date Date of manufacture Date of factory recertification R Latex Free Disposable Single patient use only Tear here to open Do...

Страница 9: ...nt of ECG leads and electrodes Symbol for the marking of electrical and electronic equipment that must be recycled Fragile handle with care Keep away from rain Keep dry This way up Stacking limit by n...

Страница 10: ...Marked by BSI 0086 per the Medical Device Directive 93 42 EEC of European Union ETL Classified by ETL Semko with respect to electric shock fire and mechanical hazards only in accordance with UL 60601...

Страница 11: ...tion training and other training as required by state province or country regulations Training on operation and use of the AED Pro Training in manual defibrillation Additional training as required by...

Страница 12: ...e contents of the box against your packing slip ENVIRONMENTAL OPERATING AND STANDBY CONDITIONS See Section 6 Technical Data Parameters Environmental Operation and Standby Conditions CAUTION Temperatur...

Страница 13: ...tery is installed but the lid is closed In this mode the AED Pro is not being used in a rescue The device will conduct its routine self tests to ensure proper operation Storage Mode When the battery i...

Страница 14: ...AED Pro without degrading its performance Note Storing the battery outside its specific range 0 50 C 30 122 F will decrease battery life WARNING Lithium Sulfur Dioxide Battery 2023681 001 9145 is Not...

Страница 15: ...y into the AED Pro as shown in the drawing until the battery snaps into place The exposed side of the battery should be flush with the outside of the AED Pro case 3 Open the lid for 5 seconds to initi...

Страница 16: ...to the rechargeable battery and ensure the yellow LED above the rechargeable battery symbol is on Charging is complete when the yellow Charge LED goes out and the four green Fuel Gauge LEDs are contin...

Страница 17: ...n plug the pad connector into the AED Pro case as shown in the drawing 4 Tuck the excess cable length in the bottom holder as shown in the drawing With the pad package completely secured to the AED Pr...

Страница 18: ...aily weekly or monthly self test determines service is required an audible beep is sounded every 30 seconds until the lid is opened or the battery power is depleted Opening and closing the lid may dea...

Страница 19: ...en the lid is first opened The text display shows the operator the identifiers for the internal code voice prompts and text prompts versions H CPR Counter During CPR a countdown timer will be displaye...

Страница 20: ...the display SETTING THE AED PRO INTERNAL CLOCK The internal clock is preset at Central Standard Time and should be reset to the correct date and local time The AED Pro will automatically adjust itself...

Страница 21: ...KIN DO NOT TOUCH PATIENT ANALYZING RHYTHM SHOCK ADVISED CHARGING STAND CLEAR PUSH BUTTON TO SHOCK PLUG IN PADS CONNECTOR SHOCK DELIVERED NOW SAFE TO TOUCH THE PATIENT START CPR 30 COMPRESSIONS 2 BREAT...

Страница 22: ...s inadvertently closed during a rescue this prompt will repeat for 15 seconds Prompt issued when the device is prepared to shock then detects a change in rhythm and therefore cancels the shock ECG Mon...

Страница 23: ...nce to initiate the manual mode Prompt issued after ALS presses the MANUAL button again to confirm Prompt issued when the MANUAL button is not pressed a second time within five seconds the device stay...

Страница 24: ...once then again to confirm the device will begin charging If the rescuer deems that the rhythm is shockable therapy can be delivered by pressing the SHOCK button Then the AED Pro reverts back to AED m...

Страница 25: ...e and Remove Pads Keep the pads connected to the AED tear the package along the dotted line and remove the pads from the package Leave the package attached to the pad wires After the prompt Peel One P...

Страница 26: ...isarm and enter CPR mode If noise is detected during analysis the AED will warn you with the prompt Analysis Interrupted Stop Patient Motion and restart the analysis This usually occurs if the patient...

Страница 27: ...conscious and breathing normally leave the pads on the patient s chest connected to the AED Make the patient as comfortable as possible and wait for Advanced Life Support ALS personnel to arrive Conti...

Страница 28: ...he manual mode is initially displayed on the screen when activated If the Medical Director has disabled this feature in ServiceLink an icon indicating NO MANUAL MODE will appear in the bottom left of...

Страница 29: ...ately sold ECG Patient Cable Kit is required to use this feature The Kit is designed for connection to ECG electrodes per AAMI or IEC color convention Once connected the AED Pro displays and evaluates...

Страница 30: ...therapy screens with the exception of the ECG MONITORING screen On the ECG MONITORING screen the Z Bar will be displayed only if the detection lead is set to Pads The Z Bar is divided into 5 sections...

Страница 31: ...other than those approved by General Electric may cause the AED Pro to function improperly during a rescue CAUTION Possible Improper AED Pro Performance Using pads that are damaged or expired may res...

Страница 32: ...THE RESCUE DATA RECORDING DATA IN INTERNAL MEMORY The AED Pro automatically stores up to 60 minutes of the latest rescue data REVIEWING THE RESCUE DATA RETRIEVING DATA FROM MEMORY 1 Open the AED Pro l...

Страница 33: ...splay and set the AED date and time Clear rescue data from the AED RescueLink is programmed with on line Help Help may be accessed by selecting Help Search for Help on from the menu bar RESCUELINK SOF...

Страница 34: ...operating system language on your computer 5 The installation program will guide you through the installation process 6 After successful installation you will be able to run RescueLink by a Selecting...

Страница 35: ...Page 34 of 53 Copyright 2007 General Electric Company 2023486 201 Rev C THIS PAGE INTENTIONALLY LEFT BLANK FOR YOUR NOTES...

Страница 36: ...Self Test completes a partial charge of the high voltage electronics current in addition to the items tested in the Daily Self Test During the Monthly Self Test the high voltage electronics are charg...

Страница 37: ...ry is low Replace with a new battery The battery power is completely depleted Replace with a new battery If STATUS INDICATOR remains RED refer to the Responder AED Pro for maintenance Call Customer Se...

Страница 38: ...emains RED 5 Open the lid and confirm that no diagnostic indicators are lit 6 Close the lid 7 Verify the STATUS INDICATOR turns GREEN Check the Integrity of the Case Examine the molded case of the AED...

Страница 39: ...ins closed for more than 15 seconds a new rescue will initiate when the lid is reopened If the remain in manual mode option has been selected in the initial set up and the AED is in manual mode when t...

Страница 40: ...PONDER AED PRO 45 ELECTROMAGNETIC COMPATIBILITY REQUIREMENTS 47 PARAMETERS OPERATION Semi Automatic shock advisory Manual AUDIBLE ALERTS Voice Prompt Maintenance Alert VISIBLE INDICATORS STATUS INDICA...

Страница 41: ...IONS Atmosphere Condition Temperature 0 C to 50 C 32 F to 122 F Humidity 5 to 95 non condensing Pressure 57kPa 4 572m 15 000ft to 103kPa 152m 500ft SHIPMENT AND TRANSPORT ENVIRONMENTAL CONDITIONS for...

Страница 42: ...he type of battery actual usage and environmental factors RECHARGEABLE BATTERY SPECIFICATIONS Battery Voltage 11 1V Chemistry Lithium ion Refer to local regulations Compatibility Responder AED Pro Mod...

Страница 43: ...a new rescue The maximum time from Analyze to Ready to Shock is 22 seconds for a new rescue AED Pro SELF TEST SEQUENCE Frequency of Self Test What is Tested Daily Weekly Monthly every 28 days Open Lid...

Страница 44: ...at or below 0 08 mV will be classified as Asystole and will not be shockable NOISE DETECTION The AED Pro will detect noise artifact in the ECG Noise could be introduced by excessive moving of the pati...

Страница 45: ...ened through a number of SVT Discriminators to classify them into VF VT or SVT Rhythms classified as SVT between the two set rates are not shockable All SVT rhythms above the rates will be classified...

Страница 46: ...sed upon a patient s thoracic impedance This customization adjusts for the unique physical differences between patients The range of impedance over which the device will deliver a shock is 25 180 Ohms...

Страница 47: ...e Energy Waveform Responder AED Pro all values are typical Phase 1 Phase 2 Patient s Impedance Ohms Voltage Volts Duration ms Voltage Volts Duration ms Energy Joules 25 1570 3 3 825 3 2 200 250 50 160...

Страница 48: ...uirements for safety 2 Collateral standard electromagnetic compatibility Requirements and tests EMISSIONS Electromagnetic Fields CISPR 11 2003 Industrial scientific and medical ISM radio frequency equ...

Страница 49: ...d for use in the electromagnetic environment specified below The customer or the user of the AED should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance leve...

Страница 50: ...r to any part of the AED including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P 8...

Страница 51: ...in which radiated RF disturbances are controlled The customer or the user of the AED can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF comm...

Страница 52: ...and one RescueLink event review CD ROM Responder AED Pro is available in more than twenty languages with others being added on a regular basis For a complete list of those available contact your GE sa...

Страница 53: ...pes one way filter mask 2019199 005 AED Wall mount storage case 2019199 006 AED Wall mount storage case with strobe light alarm 2024178 001 AED Pro 3 Lead ECG Cable Kit IEC 2024452 001 Add on Pouch fo...

Страница 54: ...USA Tel 800 558 7044 USA only Fax 800 421 6841 Canada Tel 800 668 0732 GE Medical Systems Information Technologies GmbH Munzinger Str 3 D 79111 Freiburg Germany Tel 49 761 4543 0 Fax 49 761 4543 233 R...

Страница 55: ......

Отзывы: