HygroTrace Moisture Transmitter
The Default Display
Figure 6 on page 11 shows the normal, default display of the
HygroTrace
/TFM.
•
Menu Prompt Area
- Titles, prompts, status displayed here.
•
Measurement
- 3 ½ digits (0 - 1999). Automatic decimal
placement based on range.
•
Unit Symbol
- ppb = Parts Per Billion water by volume.
•
Keypad Lock Indicator
- Unlock sequence required to modify
settings.
•
Measurement Progress Bar
-
HygroTrace
takes
~2 minutes to acquire data and complete the measurement.
The progress bar gives a visual indication of the measurement
status.
•
Warning/Error Indicator
- In the absence of an error, this
region is blank.
Errors
that effect the quality of the
measurement (e.g. heater failure) are indicated in
ALL
CAPITAL
letters.
Warnings
that indicate problems detected
with the measurement are indicated in
Mixed Case
letters.
Refer to Table 3 below.
Table 3: Warning/Error Messages
No.
Label
Type
Description
0
Status OK
----
No Errors detected.
3
Sns Ovr Rng
WARN
The calculated PPBv value is greater
than the highest calibration point,
and is an extrapolation.
12
Acquiring...
WARN
Displayed at startup while the
HygroTrace
collects data for the first
measurement.
101
HTR FAIL LO ERROR
The heater failed to reach the desired
temperature in the time allotted.
102
HTR FAIL HI ERROR The heater exceeded 200°C.
204
Sns Too Wet
WARN
The sensor is exposed to excessive
moisture conditions.