MODEL 610

              Liquid Nitrogen
Battery Operated Low Level Alarm

16380 13  Mile  Rd. • Roseville, Michigan 48066 • (586) 778-0426 • Fax (586) 778-0479

www.gordinier.com     Email: [email protected]

Содержание 610

Страница 1: ...MODEL 610 Liquid Nitrogen Battery Operated Low Level Alarm 16380 13 Mile Rd Roseville Michigan 48066 586 778 0426 Fax 586 778 0479 www gordinier com Email pat gordinier com ...

Страница 2: ...cket and Controller to Small Refrigerator 3 Low Level Probe Installation 4 Connect Probe to Controller 5 Connect Transformer to Controller 5 Option Special Order 6 Testing the Controller 7 Charging the Battery 7 Calibrating Low Level Alarm 8 Care Maintenance 9 MODEL 610 ...

Страница 3: ...arent shipping damage such as broken knobs switches or dented cases If there is any damage IMMEDIATELY contact the carrier for corrective action If an operating problem has been detected once the instrument has been properly set up contact your Gordinier Electronics Inc distributor or Gordinier Electronics Inc directly at 586 778 0426 STOP READ THIS ENTIRE INSTRUCTION MANUAL If you do not understa...

Страница 4: ...odel 610 Liquid Nitrogen Low Level Alarm is a small self contained unit with features you would expect to find only in larger costly alarms designed with a test circuit and probe protection MODEL 610 Model 610 2 Built in battery charger Continuous alarm 5 days on full battery charge Universal 9 volt AC Wall Transformer OPTION 220 volt Open or shorted probe protection Push to test button battery an...

Страница 5: ...ow this controller to come in contact with Liquid Nitrogen vapors Model 610 3 MODEL 610 Secure bracket to handle on storage unit with existing screw Remove Velcro Film from strip 2 places and position controller onto bracket Screw Velcro Film Bracket Handle Handle Controller ...

Страница 6: ...vel A minimum of 2 to 3 inches is suggested this should allow for several days of safety margin The sensor bulb is approximately 1 4 from end of probe Note Consult tank manufacturer for data on Day of holding time per inch of liquid nitrogen level Sensor Probe Liquid Nitrogen Sensor Probe Connector Controller End of Sensor Probe When proper depth is determined gently bend the probe wires at the ne...

Страница 7: ...ed to a properly grounded AC outlet Unfortunately some of the three 3 prong outlets are not properly grounded Check to make sure that the outlet to be used is connected to a known ground BEFORE plugging in the AC power cord Proper AC Outlet 9 Volt Transformer 9 Volts AC Low Sensor Probe 9 Volt Transmittor Connector B Plug the Probe A into the Controller as shown above Plug the 9 Volt AC Transforme...

Страница 8: ...ion to Controller Special Order Lid inserts directly into dewar Low Level Alarm Probe OPTION Special Order Lid WARNING Do Not allow this controller to come in contact with Liquid Nitrogen vapors MODEL 610 Model 610 6 Special Order Lid ...

Страница 9: ...tton The audible signal will sound for 5 days contiuously depending on the charge level of the batteries The audible signal can be silenced by correcting the low liquid nitrogen level problem or turning the battery switch to Off In the event that the single thermistor should open or break a tone will alert the operators The battery and alarm horn can be tested at any time When the batteries begin ...

Страница 10: ...juster clockwise or counter clockwise until the tone begins to sound Slowly turn the adjuster until the tone sound just disappears Release the Cal toggle switch to its battery Off position Press Cal toggle switch to its full upright position or battery On Press the Push to Test button to assure the battery power is turned on and calibrated ADJUST WITH SCREW DRIVER ON OFF CAL TOGGLE SWITCH PROBE TE...

Страница 11: ...emperature environ ments Temperature extremes can shorten the life of electronic devices damage batteries and distort or melt plastic parts Keep the controller away from dust and dirt which can cause premature wear of parts Wipe the controller with a damp cloth occasionally to keep it looking new Do not use harsh chemicals cleaning solvents or strong detergents to clean controller The following su...